CALIBRATION OF A LOW COST ELECTRICAL IMPEDANCE TOMOGRAPHY SYSTEM WITH A RING OF RESISTORS AND A KCL SOLUTION
Submission Author:
Bruno Furtado de Moura , GO
Co-Authors:
Bruno Furtado de Moura, francisco sepulveda, Jorge Luis Jorge Acevedo, Wellington Betencurte da Silva, Rogerio Ramos, Marcio Ferreira Martins
Presenter: Bruno Furtado de Moura
doi://10.26678/ABCM.COBEM2019.COB2019-1535
Abstract
Electrical impedance tomography (EIT) is widely to determine the flow pattern of multiphase flow. This method presents a technical advantage over others, as it is a non-intrusive and radiation-free technique. The method consists of setting electrodes in the boundary of a vessel, obtaining information about the voltages produced by the injected current. The information produced by the acquisition system is used to solve an inverse problem. This combination allows estimating the distribution of conductivity within the domain. In this work, a calibration through a ring of resistors and KCl solution vessel in order to diminish the influence of the systematic error. The random error and the signal-to-noise ratio is also evaluated for this system.
Keywords
Electrical Impedance Tomography, Acquisition System, calibration